SAS Sampler

The SAS Microbial Air Sampler is a low cost solution for companies searching to integrate viable air sampling into their isolator, RABS or pharmaceutical filling line.
  • Stainless Steel 304 Construction
  • 100lpm Controlled Flow Rate
  • Vaporised Hydrogen Peroxide Tolerant
  • Autoclave Tolerant Head
A dedicated VF8023 controller module drives each Air Sampler and can be used independently or in conjunction with enVigil software. It can be set up for continuous or intermittent sampling of 1 cubic metre. With intermittent sampling it provides a programmable number of shorter samples spread over a longer period to make up the 1 cubic metre volume. Intermittent sampling is often used to reduce the number of  interventions during a production process.
VF8023 Control Module with SAS Head
Air Sampler for integration into filling line monitoring
The Air Samplers come in a variety of versions with a large number of permutations allowing you to choose the spare part appropriate to your installation. The basic parts consist of a perforated head that allows air to be drawn in over the sample plate and a body (base) that incorporates the motorised impeller that draws the air in. The base is available in two mounting versions: a pedestal base for free standing locations or a Tri-Clamp base for integration within machines. The pedestal base uses a Lemo style connector for quick connection/disconnection, while the Tri-Clamp base has an integral cable.
A pre-formed cable assembly is available for the pedestal base, while the Tri-Clamp base generally requires a custom cable to be completed on-site. The head and base types must be matched and are available in two types: standard screw head and bayonet head for easier removal. A protective cover is also available which is suitable for both screw and bayonet heads. It should be noted that the flow rate itself is determined by the associated VF8023 Air Sampler controller, not the Air Sampler components.
Cleanroom Management International - SAS Sampler
Monitoring
Viable Monitoring for Qube Isolators and filling lines
  • Air samplers can be driven by monitoring system
  • 21 CFR part 11 Compliance
  • Available in Pedestal and Tri-clamp version
  • Available in Bayonet and Screw head
  • Head and Base are made of 304 stainless steel
  • No external vacuum required