MET ONE portable 3400-series
The METONE 3400 Series was designed for a broad range of cleanroom applications like particle counting in an aseptic production environment, process trouble shooting and classification/qualification measurements according to ISO 14644-1.
Robust, Stainless Steel Enclosure
The MET ONE 3400 series is a very robust and reliable portable particle counter with a full stainless steel enclosure and stands for accurate and reproducible particle counting according to ISO 21501-4 all in compliance with the EU-GMP Annex 1. The data integrity has been further improved with the optional paperless-pdf functionality to comply to the FDA ALCOA-Guidance. The paperless pdf functionality eliminates data errors, saves time and costs while processing raw data.
Cleanroom Applications
The MET ONE 3400 Series is a portable air particle counter, easy to use and reliable solution for a wide variety of cleanroom applications.
Industry’s first embedded secure data transfer for 21 CFR Part 11 compliance
Sample plan wizard for 14644 & Annex 1 compliance
Supports ISO 21501-4 compliance for unit-to-unit accuracy & reproducibility
One-touch sampling & location ID
Embedded location & sample recipe management
Dual hot-swappable batteries for continuous operation
Software & test wizards for easy error-free configuration
Open communications architecture for seamless facility integration & data management
Upload data to Facility Monitoring System (FMS)
MET ONE 3400 Series is easy to use and easy to integrate for any facility or environmental monitoring programme. It is easy to setup the MET ONE 3400 for routine sampling through the use of software wizards that guide the analyst to configure the instrument for ISO 14644:2015, EU GMP Annex 1 and other common regulatory sampling requirements, the 3400-series make use of Long Life Laser Diode Technology.
The 3400-series can be used for environmental monitoring of grade C and D area’s or for real-time monitoring applications in grade A and B environments. The data obtained in the grade C and D environment can be uploaded to the Envigil Monitoring System by means of a download socket in the cleanroom area.
Application 3400-series portable particle counter
Pharmaceutical Cleanroom Monitoring
Particle Monitoring in micro-electronics industry
Compressed Gas Sampling ISO-8573
HEPA filter leak detection
Microelectronics Industry
Process troubleshooting
ISO14644-1 Particle Counting